Table of Contents
On-site experiments
On-site safety protocols
As per beamline policy, it is your
responsibility to know and follow all safety protocols applicable
to your experiment, samples and equipment you bring to the SSRL.
Please note the SSRL has policies concerning the use and disposal of
propane and ethane,liquid nitrogen,
heavy
metals and compressed gases. Please follow
all the applicable rules.
Safety in the experimental hutch
To minimize the possibility of accidents while working inside the
hutch, there are restrictions on the motors that can be moved from
certain locations.
The hutches are equipped with oxygen deficiency sensors. An alarm will go
off if the oxygen content in the hutch falls below 19.5%. This may
happen while the nitrogen dewar is being refilled. Do not enter
the hutch or open the door if the oxygen alarm is sounding!. If the
duty operator arrives to investigate the oxygen alarm, explain that
the alarm will be automatically reset once the dewar has been filled. If the
alarm does not subside within a few minutes after the nitrogen filling
is finished, call support staff.
Mounting samples
The following instructions describe how to mount samples manually. For automated sample mounting, please refer to the SAM robot use instructions.
- An X-terminal monitor is available inside the hutch. You can log
in at this terminal and start Blu-Ice (see the
Blu-Ice
documentation). To be able to
mount samples comfortably, the detector should
be moved to about 400-500 mm and the beamstop to 40 mm.
- On BL12-2, make sure that you retract the backlight screen for
the On-Axis camera (from the Sample Video
interface
in Blu-Ice)
Mounting samples in the cryostream.
- The 4LD filling dewar can be used for dispensing
liquid nitrogen. It can be filled up at a filling station. The main
filling station is located in building 120 near beamline 9; other
filling dewars are available in different locations of the
experimental floor.
Figure 3:
Dewars available at the beamline. The two white Nalgene
dewars can be used for short term cassette, pucks or samples storage and
transport; the 4LD dewar can be used for dispensing liquid
nitrogen.
|
- Cryo-tools are available on the tool
board.
- A microscope is available for mounting crystals and
flash-cooling directly in the cold stream. The microscope can be
placed on a small table on top of the dewar inside the hutch. The
table (usually located inside the experiment hutch) can be installed as shown
below. Use the plastic step stored inside the
experiment hutch to reach the microscope table.
Figure 4:
Beamline microscope
|
Figure 5:
Mounting the microscope table in the experiment
hutch
|
- Roughly align the goniometer for your samples by mounting
an empty pin of the same length on the goniometer head and centering
it as described in the
Blu-Ice
documentation.
- Verify that you can easily insert the tongs to mount or remove the
sample pin.
- Check that the centered loop is in the
center of the cryostream. If it is not, contact staff.
- Check that the temperature of the cryo-cooler, displayed on the
cryo-controller, is about
100 K. Occasionally the temperature may be a few degrees higher. This
means that the unit will soon need maintenance, but it should not affect
the sample. Do not try to adjust the flow rates as this can cause icing.
Figure 6:
Cryo-controller
|
- If the crystal has snow flakes, it can be cleaned by pouring a
small amount of liquid nitrogen (use a cryo-vial) over it. Protect the
sample camera by covering it with a clean light
object (for example, the light blue foam covers for the small dewars).
Figure 7:
The sample camera
|
The Oxford cryostream can be used to collect data at temperatures
other than 100 K. The temperature can be changed at the
cryostream controller outside the hutch (
you need to ask the support staff to enable manual control of the cryojet
before being able to do this):
- On the left hand side of the controller, under the temperature display,
there is a black button labeled "set". Pressing this button will
display the set temperature. Pressing the "raise" red button to the
right of the display while pressing the "set" button will increase
the set temperature.
- Once the set temperature has reached the desired value, release
the buttons. Monitor the temperature display to find out when the
system reaches the new set temperature.
- In some cases you might have to adjust the translation of the cold
stream nozzle to accommodate the capillary or plastic sleeve. Please
contact staff in this case.
In the case that the overhead lamp does not provide adequate
illumination, the side light source can be used to better view
the sample. Use the Blu-Ice
interface
to turn the light on and adjust the intensity of the light
until you obtain a clear view of the
sample. When using automated centering of the loop, the software will
temporarily turn off the side light source. The software will also
turn off the light if the software has been idle for some time.
On BL12-2, make sure to insert the backlight screen to be able to
see small samples using the on-axis camera.
On some beamlines, a Visex
microscope
is available to detect the
crystal. The alternative, is to use low dose X-ray rastering
to find
the sample.
Checking out
- Make sure that all data are backed up or transfered to your home
computer. Read the instructions for using the available
backup
facilities.
- Return all the items you have borrowed to the toolboard; if the
small open dewars still contain liquid nitrogen you may leave them
to empty and dry off on the table (as far away from the edge as
possible).
- Throw away all garbage. Recycle white paper and printouts in the
blue recycling bins near the beamline printer. Put used sharps and
glass in the sharps boxes provided. Disposal of hazardous material
should be arranged with the safety officer (Matt Padilla, ext. 3861).
- Log in to the User Research Administration
Database
and
fill out the end of run summary. Be as specific as possible when
reporting problems at the beamline. For problems requiring prompt
attention, you can also send an e-mail to your support staff contact
person.
- If you finish, or know in advance that you will finish
your experiment more than two hours early, call
your contact person on the beamline phone or, outside normal working
hours, send them an e-mail.
- Users may leave an experiment running at the end of their run,
as long as it finishes before 11 am. Leave a note or e-mail support
staff and leave a phone number where you can be reached.
Experiments running unattended after 11 am may be stopped, unless
previously arranged with support staff.
|